Power Reference Level Test

Performance Tests

Power Reference Level Test

Power Reference Level Test

Electrical

Performance

Conditions

Characteristics

Limits

 

 

 

 

 

 

 

Power reference

1 mW

Internal 50 MHz oscillator

 

 

factory set to ±0.4% traceable to

 

 

National Physical Laboratory

 

 

(NPL), UK.

 

 

 

Power reference

±0.1%

Guardbanded Test Line Limits.

accuracy setting

 

 

 

 

 

Power reference

±0.5%

Worst case power meter

accuracy after

(23 ±3oC)

accuracy, taking into account

2 years

 

the possibility of oscillator drift

 

 

and operation at 55 degrees.

 

 

 

 

Description

 

This test verifies the power level accuracy of the internal 50 MHz

 

oscillator—also called the power reference oscillator. The power reference

 

oscillator output is factory adjusted to 1 mW ±0.4%. This accuracy

 

includes a performance limit of 0.1% and a system measurement

 

uncertainty figure of 0.3% (traceable to the National Physical Laboratory

 

(NPL), UK). Refer to Appendix B for the measurement uncertainty of the

 

system.

 

 

Note

The setting of the oscillator output accuracy is determined from the

 

system measurement uncertainty plus any chosen test line limit.

 

For example, if a test line limit of 1 mW ±0.2% was chosen while

 

maintaining a system measurement uncertainty of ±0.3%, the overall

 

accuracy of the test setup for the 1 mW power output becomes

 

0.3% + 0.2%, giving a setting accuracy of 0.5%.

 

 

Agilent E4418B/E4419B Service Guide

2-15