Performance Tests

 

Instrument Accuracy Test

 

Procedure

 

The following procedure should be performed for the Instrument Accuracy

 

Test.

 

 

Note

The procedure details the key presses required on the

 

Agilent E4418B. For the Agilent E4419B the equivalent key

 

presses should be performed on both channels.

 

 

 

1. Connect the equipment as shown in Figure 2-3.

 

2. Unplug the range calibrator's power cord. Due to a ground loop

 

design problem with the 11683A option H01 range calibrator, it

 

will not operate properly unless ground loops are eliminated. The

 

instrument will continue to operate with no power applied, and

 

disconnecting the power cord eliminates the problem.

 

 

WARNING

There is a small risk of mild electrical shock from the

 

chassis of the 11683A option H01 range calibrator unless

 

the power meter is connected.

 

 

 

3. Figure 2-3 shows the HI/LO outputs of the DC calibrator

 

connected to the DC reference input of the range calibrator.

 

Ensure that the Guard and LO output of the DC calibrator are

 

floating, and are not connected to the earth ground.

 

4. Switch the power meter on.

5.Press PreseLocal t

6.Press dBm/W

7.Press MeasSetup

to 4.

then Confirm .

, W for a reading in watts.

on the power meter, and set the Resolution value

8.Set the range calibrator as follows:

• RANGE: 100 mW

• POLARITY: NORMAL

• FUNCTION: STANDBY

• REF SWITCH: EXT

• LINE: OFF

2-12

Agilent E4418B/E4419B Service Guide