Agilent Technologies e4419b, e4418b manual Performance Test Record

Models: e4419b e4418b

1 144
Download 144 pages 36.25 Kb
Page 57
Image 57
Performance Test Record

Performance Tests

Performance Test Record

Performance Test Record

Model Agilent E4418B/E4419B Power Meter

Tested by ___________________________________________

Serial Number _______________________________________

Date _______________

Table 2-7: Performance Test Record

 

Min

Channel A

Channel B

Max

Test

Actual

Actual

Result

Result

 

Result

Result

 

 

 

 

 

 

 

 

 

 

 

 

 

Zero Test

−76.40 nW

 

 

+76.40 nW

 

 

 

 

 

Instrument Accuracy

 

 

 

 

3.1623 ∝W

3.0977 ∝W

___________

___________

3.2269 ∝W

10 ∝W

9.904 ∝W

___________

___________

10.096 ∝W

31.6228 ∝W

31.428 ∝W

___________

___________

31.819 ∝W

100 ∝W

99.49 ∝W

___________

___________

100.51 ∝W

316.227 ∝W

314.72 ∝W

___________

___________

317.74 ∝W

1 mW

0.9954 mW

___________

___________

1.0047 mW

3.1623 mW

3.1477 mW

___________

___________

3.1769 mW

10 mW

9.954 mW

___________

___________

10.046 mW

31.6228 mW

31.477 mW

___________

___________

31.769 mW

100 mW

99.54 mW

___________

___________

100.46 mW

 

 

 

 

 

Power Reference

 

 

 

 

Pmeas

0.999 mW

___________

___________

1.001 mW

Interface Test

Pass/Fail

___________

___________

Pass/Fail

 

 

 

 

 

Functional Test

 

 

 

 

Low Range, 30 dBm pad

Pass/Fail

___________

___________

Pass/Fail

Low Range, 30 dBm pad

Pass/Fail

___________

___________

Pass/Fail

Low Range, no pad

Pass/Fail

___________

___________

Pass/Fail

Upper Range, no pad

Pass/Fail

___________

___________

Pass/Fail

 

 

 

 

 

Agilent E4418B/E4419B Service Guide

2-23

Page 57
Image 57
Agilent Technologies e4419b, e4418b manual Performance Test Record