Performance Tests
Instrument Accuracy Test
Note | When switching the range calibrator to STANDBY, allow enough |
| |||||||
| time for the range calibrator to settle to its zero value before |
| |||||||
| attempting to zero the power meter. This settling would appear on | ||||||||
| the power meter display as downward drift. When the drift has |
| |||||||
| reached minimum, (typically less than 60 seconds), the range |
| |||||||
| calibrator is settled. |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 6. On the power meter press | System | , |
|
| , |
|
| , |
|
| Input Settings |
|
| |||||
|
| More |
| ||||||
| Inputs |
|
| ||||||
|
|
| |||||||
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
Filter , Filter On , Mode Man , Length , and set the filter length to 512 in the pop up window then press Enter .
7.On the power meter press ZerCalo , Zero . Wait approximately 10 seconds for the wait symbol to disappear. Verify that the display reads 0 ±0.05 ∝W.
8.Set the range calibrator’s FUNCTION switch to CALIBRATE.
9.Set the range calibrator’s RANGE switch to 1 mW.
10. Press
Cal to calibrate the power meter.
11.Set the range calibrator’s RANGE switch to the positions shown in Table