Adjustments

 

 

 

Power Reference Oscillator Frequency Adjustment

 

 

 

 

 

Power Reference Oscillator Frequency Adjustment

 

 

 

Note

 

Adjustment of the power reference oscillator frequency may also

 

 

affect the output level of the oscillator. Therefore, after the

 

 

frequency is adjusted to 50.0 ±0.5 MHz, the output level should be

 

 

checked as described in “Power Reference Oscillator Level

 

 

Adjustment”, on page 3-6.

 

 

 

The power reference oscillator frequency is a nominal

 

 

specification.

 

 

 

 

 

 

 

Description

 

 

 

Variable inductor A2L9 is adjusted to set the power reference oscillator

 

 

output frequency to 50.0 ±0.5 MHz. This frequency is a supplemental

 

 

characteristic.

 

 

 

Power Meter

Frequency Counter

POWER

REF

Figure 3-1: Adjustment Setup

Equipment

Frequency Counter

Agilent 53132A

Procedure

1.Ensure that the power meter has been powered on for at least 30 minutes before making any measurements.

2.Connect the equipment as shown in Figure 3-1. Set up the counter to measure frequency.

3-4

Agilent E4418B/E4419B Service Guide

Page 62
Image 62
Agilent Technologies e4418b, e4419b manual Power Reference Oscillator Frequency Adjustment