Figure 2-3: Instrument Accuracy (Average Path) Test Setup

 

Performance Tests

 

Instrument Accuracy Test

 

Test Procedure 2

 

 

Note

This test procedure makes use of a 11683A Range Calibrator fitted

 

with option H01. The measurement uncertainty of this test

 

procedure is significantly lower than the previously described

 

method (refer to Appendix A for a description of the measurement

 

uncertainty analysis). As such, this test procedure should be used

 

to ensure the power meter meets it's published specification.

 

 

Equipment

Range Calibrator: Agilent 11683A Option H01

Power Sensor Cable: Agilent 11730A

DC Range Calibrator: Datron 4000A

Test Setup

DC Calibrator

Power Meter

CHANNEL A

Range

 

 

Power sensor cable Calibrator

Figure 2-3: Instrument Accuracy (Average Path) Test Setup

Agilent E4418B/E4419B Service Guide

2-11

Page 45
Image 45
Agilent Technologies e4419b, e4418b manual Test Procedure, Equipment, 3 Instrument Accuracy Average Path Test Setup