| Performance Tests |
| Instrument Accuracy Test |
| Test Procedure 2 |
|
|
Note | This test procedure makes use of a 11683A Range Calibrator fitted |
| with option H01. The measurement uncertainty of this test |
| procedure is significantly lower than the previously described |
| method (refer to Appendix A for a description of the measurement |
| uncertainty analysis). As such, this test procedure should be used |
| to ensure the power meter meets it's published specification. |
|
|
Equipment
•Range Calibrator: Agilent 11683A Option H01
•Power Sensor Cable: Agilent 11730A
•DC Range Calibrator: Datron 4000A
Test Setup
DC Calibrator
Power Meter
CHANNEL A | Range |
| |
| Power sensor cable Calibrator |
Figure 2-3: Instrument Accuracy (Average Path) Test Setup
Agilent E4418B/E4419B Service Guide |