Specifications
Power Meter Supplemental Characteristics
Settling Time
0 to 99% settled readings over the
For Agilent 8480 series power sensors
Manual filter, 10 dB decreasing power step:
Table 1-4: Settling Time
Number of |
| 1 | 2 | 4 | 8 | 16 | 32 | 64 | 128 | 256 | 512 | 1024 |
Averages |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Settling Time (s) |
| 0.15 | 0.2 | 0.3 | 0.5 | 1.1 | 1.9 | 3.4 | 6.6 | 13 | 27 | 57 |
(Normal Mode) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Response Time (s) |
| 0.15 | 0.18 | 0.22 | 0.35 | 0.55 | 1.1 | 1.9 | 3.5 | 6.9 | 14.5 | 33 |
(x2 mode) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Auto filter, default resolution, 10 dB decreasing power step, normal and x2 speed modes:
| Normal Mode |
|
|
|
| x2 Mode |
|
|
|
| ||||||
| Maximum dBm |
|
|
| Maximum dBm |
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 150 ms |
| 20 dB |
|
|
| 150 ms |
|
| 20 dB |
|
| ||||
|
|
|
|
|
|
|
|
| ||||||||
Typical |
|
|
|
|
|
|
| Typical |
|
|
|
|
|
|
|
|
|
|
|
|
| Power |
|
|
|
|
|
| Power | ||||
|
|
|
|
|
|
|
|
|
|
| ||||||
Settling | 200 ms |
| 10 dB | Sensor | Settling | 180 ms |
|
| 10 dB | Sensor | ||||||
Times |
| Dynamic | Times |
|
| Dynamic | ||||||||||
| 500 ms |
| 10 dB | Range |
| 350 ms |
|
| 10 dB | Range | ||||||
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
| ||||||||
| 6.6 s |
| 10 dB |
|
|
| 3.5 s |
|
| 10 dB |
|
| ||||
|
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Minimum | dBm |
|
|
|
| Minimum | dBm |
|
|
Agilent E4418B/E4419B Service Guide |