Agilent Technologies e4419b For Agilent 8480 series power sensors, Settling Time s, x2 mode

Models: e4419b e4418b

1 144
Download 144 pages 36.25 Kb
Page 25
Image 25
Settling Time

Specifications

Power Meter Supplemental Characteristics

Settling Time

0 to 99% settled readings over the GP-IB.

For Agilent 8480 series power sensors

Manual filter, 10 dB decreasing power step:

Table 1-4: Settling Time

Number of

 

1

2

4

8

16

32

64

128

256

512

1024

Averages

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Settling Time (s)

 

0.15

0.2

0.3

0.5

1.1

1.9

3.4

6.6

13

27

57

(Normal Mode)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Response Time (s)

 

0.15

0.18

0.22

0.35

0.55

1.1

1.9

3.5

6.9

14.5

33

(x2 mode)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Auto filter, default resolution, 10 dB decreasing power step, normal and x2 speed modes:

 

Normal Mode

 

 

 

 

x2 Mode

 

 

 

 

 

Maximum dBm

 

 

 

Maximum dBm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

150 ms

 

20 dB

 

 

 

150 ms

 

 

20 dB

 

 

 

 

 

 

 

 

 

 

 

Typical

 

 

 

 

 

 

 

Typical

 

 

 

 

 

 

 

 

 

 

 

 

 

Power

 

 

 

 

 

 

Power

 

 

 

 

 

 

 

 

 

 

 

Settling

200 ms

 

10 dB

Sensor

Settling

180 ms

 

 

10 dB

Sensor

Times

 

Dynamic

Times

 

 

Dynamic

 

500 ms

 

10 dB

Range

 

350 ms

 

 

10 dB

Range

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6.6 s

 

10 dB

 

 

 

3.5 s

 

 

10 dB

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Minimum

dBm

 

 

 

 

Minimum

dBm

 

 

Agilent E4418B/E4419B Service Guide

1-9

Page 25
Image 25
Agilent Technologies e4419b manual For Agilent 8480 series power sensors, Settling Time s, Normal Mode, Response Time s