Agilent Technologies e4419b Instrument Accuracy Test, Test Procedure, Specification, Description

Models: e4419b e4418b

1 144
Download 144 pages 36.25 Kb
Page 41
Image 41
Instrument Accuracy Test

Performance Tests

Instrument Accuracy Test

Instrument Accuracy Test

Specification

Electrical Characteristics

Performance Limits

 

 

 

 

Accuracy

±0.5% or ±0.02 dB 1

1.This performance limit does not include the corresponding sensor power linearity specification.

Description

The power meter accuracy is verified for various power inputs. There are two different methods described here that can be used to perform this test, although one of these methods has a relatively high measurement uncertainty values associated with it.

 

Test Procedure 1

 

 

Note

The measurement uncertainty of this test procedure is relatively

 

high at power levels of 30uW and below, as it uses a standard

 

11683A Range Calibrator. As such, this test procedure should be

 

used for indication only.

 

 

Equipment

Range Calibrator

Agilent 11683A

Power Sensor Cable

Agilent 11730A

Agilent E4418B/E4419B Service Guide

2-7

Page 41
Image 41
Agilent Technologies e4419b, e4418b manual Instrument Accuracy Test, Test Procedure, Specification, Description, Equipment