Performance Tests
Instrument Accuracy Test
Instrument Accuracy Test
Specification
Electrical Characteristics | Performance Limits |
|
|
|
|
Accuracy | ±0.5% or ±0.02 dB 1 |
1.This performance limit does not include the corresponding sensor power linearity specification.
Description
The power meter accuracy is verified for various power inputs. There are two different methods described here that can be used to perform this test, although one of these methods has a relatively high measurement uncertainty values associated with it.
| Test Procedure 1 |
|
|
Note | The measurement uncertainty of this test procedure is relatively |
| high at power levels of 30uW and below, as it uses a standard |
| 11683A Range Calibrator. As such, this test procedure should be |
| used for indication only. |
|
|
Equipment
Range Calibrator | Agilent 11683A |
Power Sensor Cable | Agilent 11730A |
Agilent E4418B/E4419B Service Guide |