Display Brightness and Contrast Adjustment

Adjustments

Display Brightness and Contrast Adjustment

Display Brightness and Contrast Adjustment

 

Introduction

 

The following procedure should be performed whenever a front panel

 

assembly or processor assembly are replaced.

 

The brightness is controlled automatically after executing the

 

 

 

softkey, located under the

 

softkey menu.

 

 

Set Brightness

Service

 

 

 

 

 

 

Note

 

The contrast adjustment is subjective and varies according to

 

 

individual user requirements.

 

 

 

 

 

 

Procedure

1.Access the Display softkey menu and use the IntroductionSet BrightnessServiceManual background softkey to

increase the contrast, or the Manual backgroundManual backgroundManual background softkey to decrease the contrast of the display.

2.When the desired display contrast level is obtained, press the Set Contrast softkey, located under the Service softkey menu.