Agilent Technologies e4419b, e4418b manual + Input, Thermistor Mount, Vcomp

Models: e4419b e4418b

1 144
Download 144 pages 36.25 Kb
Page 65
Image 65
+ INPUT

Adjustments

Power Reference Oscillator Level Adjustment

Note

The power meter may be returned to the nearest Agilent Technologies

 

office to have the power reference oscillator checked and/or adjusted.

 

 

Power Meter Power

ref

Agilent 432A

Power Meter Digital

Voltmeter

VRF

+ INPUT

 

Vcomp

INPUT

 

Thermistor

Mount

Figure 3-3: Adjustment Setup

For this adjustment the following mathematical assumptions are made:

Equation 1:

2Vcomp(V1 V0) + V02 V12

P = ----------------------------------------------------------------------------------

meas 4R(CalibrationFactor)

can be manipulated to give the following:

Equation 2:

V1 = Vcomp + V0 VcompINPUTVcomp2 (10)3(4R)(CalibrationFactor)

by using the following assumptions:

V02 V12 = (V1V0)2

The error in doing this is:

(V12 + V02 2V1V0) (V02 V12) = 2V02 + 2V1V0 = 2V0(V1 V0)

So if 2V0(V1 V0) « 2Vcomp( V1 V0) , that is, V0 « Vcomp then the error can be ignored. In practice Vcomp ~ 4Volts and V0 < 400∝V . The error is less than 0.01%.

Agilent E4418B/E4419B Service Guide

3-7

Page 65
Image 65
Agilent Technologies e4419b, e4418b manual + Input, Thermistor Mount, Vcomp