Adjustments
Power Reference Oscillator Level Adjustment
14.Remove the power meter’s cover and adjust A2R90 until the DVM indicates the calculated value of V1. Refer to Figure
A2R90
BT1
A2 Assembly | A2L9 |
A2J13 J12
Front Panel
Figure 3-4: A2R90 Adjustment Location
Measurement Uncertainty
Refer to Appendix B “Measurement Uncertainty Analysis – Power Reference Level Test”
Agilent E4418B/E4419B Service Guide |