Agilent Technologies e4418b, e4419b manual Measurement Uncertainty, 4 A2R90 Adjustment Location

Models: e4419b e4418b

1 144
Download 144 pages 36.25 Kb
Page 68
Image 68
Figure 3-4: A2R90 Adjustment Location

Adjustments

Power Reference Oscillator Level Adjustment

14.Remove the power meter’s cover and adjust A2R90 until the DVM indicates the calculated value of V1. Refer to Figure 3-4 for the position of A2R90:

A2R90

BT1

A2 Assembly

A2L9

A2J13 J12

Front Panel

Figure 3-4: A2R90 Adjustment Location

Measurement Uncertainty

Refer to Appendix B “Measurement Uncertainty Analysis – Power Reference Level Test”

3-10

Agilent E4418B/E4419B Service Guide

Page 68
Image 68
Agilent Technologies e4418b, e4419b manual Measurement Uncertainty, 4 A2R90 Adjustment Location