Measurement Uncertainty Analysis – Instrument Accuracy Test
Measurement Introduction
Standard Uncertainty u(xi):
Table
Table A-1: Standard Uncertainty Results
Power Level | Standard Deviation of 30 | Conversion to Linear Units | |
Readings | Standard Uncertainty | ||
| |||
|
|
| |
0.0132dB | 0.0097uW | ||
|
|
| |
0.0038dB | 0.009uW | ||
|
|
| |
0.0013dB | 0.0095uW | ||
|
|
| |
0.0015dB | 0.035uW | ||
|
|
| |
0.0004dB | 0.03uW | ||
|
|
| |
u(X(0)) 1mW | 0.0003dB | 0.00007mW | |
|
|
| |
u(X(5)) 3.16227mW | 0.0003dB | 0.00022mW | |
|
|
| |
u(X(10)) 10mW | 0.0003dB | 0.000690mW | |
|
|
| |
u(X(15) 31.62277mW | 0.0003dB | 0.0022mW | |
|
|
| |
u(X(20)) 100mW | 0.0004dB | 0.0092mW | |
|
|
|
The distributions are all normal.
3458A DMM Uncertainty:
The uncertainty of the voltage measured by the 3458A is:
Error = (ppm of reading + ppm of range) + (factory traceability) + (temperature coefficient)
Conditions: 1 Year Calibration and temperature variance
Reference: 3458A Operating Manual Appendix A
Factory Traceability: From 6ppm to 0.5ppm Standard Uncertainty depending on range.