Performance Tests
Introduction
Performance Test Record
Results of the performance tests may be tabulated in Table
Performance Tests
The performance tests given in this chapter are suitable for incoming inspection or preventive maintenance. During any performance test, all shields and connecting hardware must be in place.
The tests are designed to verify published power meter specifications. Perform the tests in the order given and record the data in Table
Calibration Cycle
This power meter requires periodic verification of performance to ensure that it is operating within specified tolerances. The performance tests described in this chapter should be performed once every two years. Under conditions of heavy usage or severe operating environments, the tests should be more frequent. Adjustments that may be required are described in Chapter 3.
Test Procedures
It is assumed that the person performing the following tests understands how to operate the specified test equipment. Equipment settings, other than those for the power meter, are stated in general terms. It is assumed that the person will select the proper cables, adapters, and probes required for test setups illustrated in this chapter.
Agilent E4418B/E4419B Service Guide |