SCPI Command Reference :SYSTem:DISPlay:LED

:SYSTem:DISPlay:LED

This command enables/disables the front panel Light Emitting Diodes matrix when the B2200 is in the GPIB remote mode. The front panel LED matrix will be automatically set to ON by the mode transition from GPIB remote to GPIB local.

Syntax

:SYSTem:DISPlay:LED state

 

Parameter

 

Explanation

 

 

 

 

 

state

 

ON or 1: enables the LEDs (initial setting)

 

 

 

OFF or 0: disables the LEDs

Example

OUTPUT @Agb2200;":SYST:DISP:LED 1"

 

:SYSTem:DISPlay:STRing

 

When the B2200 is in the GPIB remote mode, you can display a string (maximum

 

39 characters) on the front panel LCD. This command specifies the string and

 

displays it on the LCD.

 

The specified string data is only displayed on the front panel LCD. It is not

 

memorized. So a display-change operation will clear and delete the string data.

Syntax

:SYSTem:DISPlay:STRing 'string'

 

 

 

 

 

Parameter

 

Explanation

 

 

 

 

string

String to be displayed on the front panel LCD. Maximum

 

 

38 characters.

 

 

Numeric characters, alphabetic characters; upper case and

 

 

lower case, space, and the following symbols are available.

 

 

! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { } ^ _

 

 

 

Example

OUTPUT @Agb2200;":SYST:DISP:STR 'E:SMU1,B:SMU2,C:SMU3'"

 

This example displays E:SMU1,B:SMU2,C:SMU3 on the front panel LCD.