SCPI Command Reference [:ROUTe]:AGND:PORT

 

 

[:ROUTe]:AGND:PORT

 

 

This command specifies the input Ground Port for the specified card. For each card,

 

 

you can specify the same or different Ground Port. At *RST, the Ground Port is 12.

 

 

The ground mode is set by “[:ROUTe]:AGND[:STATe]”.

 

 

The query returns the input port number of the Ground Port.

 

 

The input ground port and a ground enabled input port cannot be assigned to the

NOTE

 

 

 

same input port.

 

 

 

If the input ground port and a couple port have been assigned to the same input port,

 

 

the ground mode and the couple mode cannot be used in parallel.

 

 

0 V output voltage source should be connected to the input Ground Port. Or set

 

 

ground_port to 12, and open the input port 12 that is the default ground port. If you

 

 

assign the input 12 as the ground port, this ground port will be internally connected

 

 

to the ground when the ground mode is set to ON.

Syntax

 

[:ROUTe]:AGND:PORT card_number,ground_port

 

 

[:ROUTe]:AGND:PORT? card_number

 

 

 

 

 

 

Parameter

Explanation

 

 

 

 

 

 

card_number

For Auto Config mode: 0 or ALL

 

 

 

For Normal Config mode: 1, 2, 3, 4, or ALL

 

 

 

ALL is not available for query.

 

 

ground_port

Input port number: 1 to 14. Or -1to disable ground

 

 

 

port.

 

 

 

 

Query response

 

port_number <newline><^END>

 

 

port_number = 1 to 14 or -1. Number set to ground_port.

Example

 

OUTPUT @Agb2200;":ROUT:AGND:PORT 0,13"

 

 

OUTPUT @Agb2200;":ROUT:AGND:PORT? 0"

 

 

ENTER @Agb2200;A

 

In this example, A will be 13.

Agilent B2200 User’s Guide, Edition 2

5-27

Page 153
Image 153
Agilent Technologies B2201A, B2200A manual ROUTeAGNDPORT