Agilent Technologies B2201A Too many relays closed. Max 52 relays/card, Bad unused port number

Models: B2200A B2201A

1 234
Download 234 pages 29.64 Kb
Page 233
Image 233

 

Error Messages

 

B2200 Specific Error Messages

 

 

Error

Error Message and Description

Number

 

 

 

3017

Too many relays closed. Max 52 relays/card.

 

Too many relays have been closed. To close new relays, open some

 

relays. Maximum 52 relays can be closed for each module.

 

 

3018

Can't change to ACONfig mode. Check card config.

 

The present card configuration of the B2200 does not allow the auto

 

configuration mode. Change the module configuration.

 

 

3019

Cannot use same port for Couple and Bias

 

The couple mode and the bias mode cannot be used in parallel when

 

a couple port and the input bias port have been assigned to the same

 

input port.

 

 

3020

Bad auto ground port number

 

Specify the output ports effective for the input ground port. The port

 

number must be 1 to 14

 

 

3021

Bad unused port number

 

Specify the input ports effective for the ground enabled input port

 

(unused port). The port number must be 1 to 9.

 

 

3022

Cannot directly specify auto ground port channel

 

Relays on the input ground port cannot be controlled directly when

 

the ground mode is ON.

 

 

3023

Cannot directly specify unused port channel

 

Relays on the ground enabled input port (unused port) cannot be

 

controlled directly when the ground mode is ON.

 

 

3024

Cannot use same port for Couple and Auto Ground

 

The couple mode and the ground mode cannot be used in parallel

 

when a couple port and the input ground port have been assigned to

 

the same input port.

 

 

3025

Cannot use same port for Unused and Auto Ground

 

The input ground port and the ground enabled input port (unused

 

port) cannot be assigned to the same input port.

 

 

 

 

Agilent B2200 User’s Guide, Edition 2

7-13

Page 233
Image 233
Agilent Technologies B2201A Too many relays closed. Max 52 relays/card, Cant change to ACONfig mode. Check card config