WARNING

CAUTION

NOTE

Installation

Requirements

For protection from electrical shock, the power cable ground must not be defeated.

Operating Environment

The Agilent B2200 must be operated within the following environmental conditions:

Temperature: 5 °C to 35 °C

Humidity: 5 % to 70 % R.H., non-condensing

Altitude: 0 m to 2,000 m

Storage and Shipping Environment

The Agilent B2200 should be stored or shipped in environments within the following limits:

Temperature: -20 °C to 70 °C

Humidity:

B2200A/B2210A:

< 80 % R.H. (at 35 °C), < 60 % R.H. (at 65 °C), non-condensing

B2201A/B2211A:

< 80 % R.H. (at 65 °C), non-condensing

Altitude: 0 m to 15,240 m

When storing the Agilent B2200

Protect the Agilent B2200 from temperature extremes to prevent condensation from forming inside the Agilent B2200. If condensation occurs, the Agilent B2200 may damage or may not satisfy the specifications. If the Agilent B2200 was placed at the high temperature and high humidity environment long time, the Agilent B2200 may degrade its performance.

When shipping the Agilent B2200

The Agilent B2200 must be packed in the certain packing materials for protection from damage when it is shipped. Cover the Agilent B2200 by using the moisture-proof and dehumidifying packing material (furnished), and pack it by using packing materials such as carton box and cushioning materials.

Agilent B2200 User’s Guide, Edition 2

2-5

Page 39
Image 39
Agilent Technologies B2201A, B2200A manual Operating Environment, Storage and Shipping Environment