Installation
Requirements
For protection from electrical shock, the power cable ground must not be defeated.
Operating Environment
The Agilent B2200 must be operated within the following environmental conditions:
•Temperature: 5 °C to 35 °C
•Humidity: 5 % to 70 % R.H.,
•Altitude: 0 m to 2,000 m
Storage and Shipping Environment
The Agilent B2200 should be stored or shipped in environments within the following limits:
•Temperature:
•Humidity:
B2200A/B2210A:
< 80 % R.H. (at 35 °C), < 60 % R.H. (at 65 °C),
B2201A/B2211A:
< 80 % R.H. (at 65 °C),
•Altitude: 0 m to 15,240 m
When storing the Agilent B2200
Protect the Agilent B2200 from temperature extremes to prevent condensation from forming inside the Agilent B2200. If condensation occurs, the Agilent B2200 may damage or may not satisfy the specifications. If the Agilent B2200 was placed at the high temperature and high humidity environment long time, the Agilent B2200 may degrade its performance.
When shipping the Agilent B2200
The Agilent B2200 must be packed in the certain packing materials for protection from damage when it is shipped. Cover the Agilent B2200 by using the