Front Panel Operation

Display Functions

LCD

The LCD displays the setup information and the status information as shown in Table 3-5. Display example is shown in Figure 3-2.

Figure 3-2

LCD Display Example

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

01

02

03

04

05

06

 

07

08

09

 

10

11

12

13

14

 

 

 

 

-

C

C

-

FREE

 

 

G

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A

 

 

 

 

 

BBM

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lock

Error

 

Bias

 

 

Gnd

 

Rule

 

 

 

Sequence

 

 

 

 

 

 

 

 

 

 

 

Shift

The first line shows the input port status, couple port, bias port, ground port, or ground enabled port.

The second line shows the instrument status and the mode status. The example of Figure 3-2 notifies you the following information.

Input 5 and 6 are a couple port.

Input 12 is the input ground port.

Input 4 and 7 are the ground enabled input port.

Channel configuration mode is auto.

Couple mode is ON.

Ground mode is ON.

Connection rule is free.

Connection sequence is break before make.