SCPI Command Reference

 

 

 

[:ROUTe]:COUPle:PORT

 

 

[:ROUTe]:COUPle:PORT

 

 

If the input bias port and a couple port have been assigned to the same input port, the

NOTE

 

 

 

bias mode and the couple mode cannot be used in parallel.

 

 

If the input ground port and a couple port have been assigned to the same input port,

 

 

the ground mode and the couple mode cannot be used in parallel.

 

 

If the ground enabled input port and a couple port have been assigned to the same

 

 

input port, the ground mode and the couple mode cannot be used in parallel.

 

 

This command is used to set up input couple ports, which are used for the Kelvin

 

 

connections. The couple mode is set by “[:ROUTe]:COUPle[:STATe]”.

 

 

For each card, you can set up the same or different input couple ports.

 

 

This command overwrites the previous couple port setting. The couple port setting

 

 

will be cleared and updated by “[:ROUTe]:COUPle:PORT:DETect”.

 

 

The query returns the odd input port number of each coupled pair.

 

 

At *RST, no input ports are coupled.

Syntax

 

[:ROUTe]:COUPle:PORT card_number,'couple_port'

 

 

[:ROUTe]:COUPle:PORT? card_number

 

 

 

 

 

 

Parameter

Explanation

 

 

 

 

 

 

card_number

For Auto Config mode: 0 or ALL

 

 

 

For Normal Config mode: 1, 2, 3, 4, or ALL

 

 

 

ALL is not available for query.

 

 

couple_port

One or more input port numbers: 1, 3, 5, 7, 9, 11 or 13.

 

 

 

Enclose by single quotation marks. Separate multiple input

 

 

 

port numbers by comma. For example: '1,5'

Query response

 

port_number{, port_number} <newline><^END>

 

 

port_number = 1 or 3 or 5 or 7 or 9 or 11 or 13

Example

 

OUTPUT @Agb2200;":ROUT:COUP:PORT ALL,'1,3'"

 

 

OUTPUT @Agb2200;":ROUT:COUP:PORT? 1"

 

 

ENTER @Agb2200;A$

In this example, A$ will be 1,3.