Front Panel Operation

Switch Control Functions

Couple Mode

Couple mode is useful for making the Kelvin connections. When the couple mode is ON, the input couple ports will be connected to the output couple ports. For the available couple ports, see Table 3-2.

For the couple port connections, the odd/even number of the input port will be connected to the odd/even number of the output port respectively. Then the even number is always the odd number plus 1.

You can independently specify the same or different couple ports and control the couple mode for each module when the configuration mode is normal.

Table 3-2

Input Couple Port Numbers and Output Couple Port Numbers

 

 

 

 

 

 

Available input

Input ports

Available output couple port No.

 

couple port No.

 

 

 

 

 

 

 

 

 

 

1

SMU 1,

2

1 and 2, 3 and 4, . . . , 11 and 12

 

 

 

 

(for normal configuration mode)

 

3

SMU 3,

4

 

1 and 2, 3 and 4, . . . , 47 and 48

 

 

 

 

 

5

SMU 5,

6

 

(for auto configuration mode)

 

 

 

 

 

7

SMU 7,

8

 

 

 

 

 

 

9

AUX 9, 10

 

 

 

 

 

 

11

AUX 11, 12

 

 

 

 

 

 

 

13

AUX 13,

14

 

 

 

(CMH, CML)

 

 

 

 

 

 

For example, if you specify the SMU input 1 for the couple port and set the couple mode ON, the input 1 will be connected to an odd number of the output port and the input 2 will connected to the next number of the output port. And if you specify to connect the input 1 to the output 10, the input 1 will be connected to the output 9 and the input 2 will be connected to the output 10.

NOTE

If the bias input port and a couple port have been assigned to the same input port, the

 

bias mode and the couple mode cannot be used in parallel.

 

If the ground input port and a couple port have been assigned to the same input port,

 

the ground mode and the couple mode cannot be used in parallel.

 

If a ground enabled input port and a couple port have been assigned to the same

 

input port, the ground mode and the couple mode cannot be used in parallel.

3-20

Agilent B2200 User’s Guide, Edition 2

Page 80
Image 80
Agilent Technologies B2200A, B2201A manual Couple Mode, Input Couple Port Numbers and Output Couple Port Numbers