Error Messages

Standard SCPI Error Messages

Error

Error Message and Description

Number

 

 

 

-148

Character data not allowed

 

A character parameter is not allowed for this position.

 

 

-150

String data error

 

An error was detected in a string parameter. This error is reported if

 

the B2200 cannot determine a more specific error -151 and -158.

-151

Invalid string data

 

An invalid string parameter data was received; for example, an END

 

message was received before the terminal quote character.

 

 

-158

String data not allowed

 

A string parameter data was received but was not allowed at this

 

point.

 

 

-160

Block data error

 

An error was detected in a block data. This error is reported if the

 

B2200 cannot determine more specific errors -161 and -168.

-161

Invalid block data

 

An invalid block data was received; for example, an END message

 

was received before the length was satisfied.

 

 

-168

Block data not allowed

 

A legal block data was received but was not allowed at this point.

 

 

-170

Expression error

 

An error was detected in an expression. This error is reported if the

 

B2200 cannot determine more specific errors -171 and -178.

-171

Invalid expression

 

The expression was invalid; for example, unmatched parentheses or

 

an illegal character.

 

 

-178

Expression data not allowed

 

An expression was received but was not allowed at this point.

 

 

7-6

Agilent B2200 User’s Guide, Edition 2

Page 226
Image 226
Agilent Technologies B2200A Character data not allowed, String data error, Invalid string data, String data not allowed