SCPI Command Reference [:ROUTe]:CONNection:SEQuence

[:ROUTe]:CONNection:SEQuence

This command specifies the connection sequence mode, which is the open/close sequence of the relays when connection route is changed from an existing connection to a new connection. The query returns the connection sequence mode of the specified card. At *RST, BBMake is selected. See “Connection Sequence” on page 3-15.

The connection sequence is used only for cards that have been set to SROUte connection rule. Also see “[:ROUTe]:CONNection:RULE”.

Syntax

[:ROUTe]:CONNection:SEQuence card_number,sequence

 

[:ROUTe]:CONNection:SEQuence? card_number

 

 

 

 

Parameter

Explanation

 

 

 

 

card_number

For Auto Config mode: 0 or ALL

 

 

For Normal Config mode: 1, 2, 3, 4, or ALL

 

 

ALL is not available for query.

 

sequence

NSEQ: No-Sequence mode

 

 

BBM: Break-Before-Make mode

 

 

MBBR: Make-Before-Break mode

Query response

NSEQ or BBM or MBBR <newline><^END>

Example

OUTPUT @Agb2200;":ROUT:CONN:SEQ ALL,MBBR"

 

OUTPUT @Agb2200;":ROUT:CONN:SEQ? 0"

 

ENTER @Agb2200;A$

 

In this example, A$ will be MBBR.