Front Panel Operation

Setup Menus

KEY

Displays the sub menu that provides the following functions.

 

EXECUTE

 

 

Displays the following message.

 

DIAG:KEY

Press any key.

Press any front panel key and confirm the LCD display. For example, if you press the Port Function key, the following message must be displayed.

DIAG:KEY [Port Function]

Continue this for all keys. After that, press the Enter key twice for the normal operation. The test result will be PASS.

 

 

If you find any wrong response, press the Exit key twice. The test result will be

 

 

FAIL.

 

 

 

Finally, press the Exit key to return to the previous menu.

 

RESULT

 

 

 

Displays the test result. Press the Exit key to return to the previous menu.

BEEPER

Displays the sub menu that provides the following functions.

 

EXECUTE

 

 

 

Starts the beeper test, and displays the following message.

 

 

DIAG:BEEP

Is beeper making 2 sounds?

 

 

Press the Enter key if the beeper works correctly. The test result will be recorded

 

 

as PASS.

 

 

 

If you find any problem, press the Exit key. The test result will be recorded as

 

 

FAIL.

 

 

 

Finally, press the Exit key to return to the previous menu.

 

RESULT

 

 

 

Displays the test result. Press the Exit key to return to the previous menu.