SCPI Command Reference

 

 

 

:SYSTem:CTYPe?

 

:SYSTem:CTYPe?

 

 

This query command returns the ID of the specified card.

Syntax

:SYSTem:CTYPe? card_number

 

 

 

 

 

Parameter

 

Explanation

 

 

 

 

 

card_number

 

For Auto Config mode: 0

 

 

 

For Normal Config mode: 1, 2, 3, or 4

Query response

AGILENT TECHNOLOGIES,model,0,revision <newline><^END>

 

model = B2210A or B2211A

 

revision = 1 or later version

 

 

If designated card is not installed or initialize failed, this command returns

 

NONE,NONE,0,0.

 

Example

OUTPUT @Agb2200;":SYST:CTYPE? 1"

 

ENTER @Agb2200;A$

 

 

:SYSTem:DISPlay:LCD

 

This command enables/disables the front panel Liquid Crystal Display when the

 

B2200 is in the GPIB remote mode. The front panel LCD will be automatically set

 

to ON by the mode transition from GPIB remote to GPIB local.

Syntax

:SYSTem:DISPlay:LCD state

 

 

 

 

Parameter

 

Explanation

 

 

 

 

 

state

 

ON or 1: enables the LCD

 

 

 

OFF or 0: disables the LCD (initial setting)

 

Even though state is OFF, some status indicators on the LCD will be effective.

Example

OUTPUT @Agb2200;":SYST:DISP:LCD 1"

Agilent B2200 User’s Guide, Edition 2

5-47

Page 173
Image 173
Agilent Technologies B2201A, B2200A manual SYSTemCTYPe?, SYSTemDISPlayLCD