VXI plug&play Driver

 

 

Driver Functions

 

agb220xa_biasChanCard

 

This function bias-enables or disables all the output ports of the specified card.

Syntax

ViStatus _VI_FUNC agb220xa_biasChanCard(ViSession vi,

 

ViInt16 disable_enable, ViInt16 bias_cardno);

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

disable_enable

Status of the card, bias enabled or bias disabled.

 

 

0 : sets bias enabled card.

 

 

1 : sets bias disabled card.

 

bias_cardno

Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5

 

 

(all card) in the normal configuration mode, or 0 (all card in the

 

 

auto configuration mode). For the configuration mode, see

 

 

agb220xa_func.

 

agb220xa_biasChanList

 

This function bias-enables or disables all the output ports specified by the

 

biaschan_list.

 

 

The parameter “biaschan_list” is an array of integers with each integer representing

 

one channel. The last number of the “biaschan_list” should be “0” (numeric zero) to

 

identify the end of the list. The maximum number of channels that can be specified

 

by the list is 100.

 

Syntax

ViStatus _VI_FUNC agb220xa_biasChanList(ViSession vi,

 

ViInt16 biaschan_disen, ViInt32_VI_FAR biaschan_list[ ] );

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

biaschan_disen

Status of the port, bias enabled or bias disabled.

 

 

0 : sets bias enabled port.

 

 

1 : sets bias disabled port.

 

biaschan_list[ ]

Channel numbers. 5 digits integer. ABCDE. where A: card

 

 

number, BC: input port number, DE: output port number.

 

 

Top zero(s) can be ignored. For example, if A=0, BC=01, and

 

 

DE=01, channel number should be 101 instead of 00101.

Agilent B2200 User’s Guide, Edition 2

6-7

Page 193
Image 193
Agilent Technologies B2201A, B2200A manual Agb220xabiasChanCard, Agb220xabiasChanList