Installation

Output Connections

Table 2-4

 

 

Connection to the DUT Interface

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Kelvin connections

 

 

 

 

 

non-Kelvin connections

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This connection is available only for the

 

Following connection is for the Kelvin

 

 

Kelvin connectors.

 

 

 

 

 

connectors. For the triaxial connectors,

 

 

This connection can cancel effects of cable

ignore SENSE terminal, and make

 

 

connection only for FORCE terminal.

 

 

resistance by connect sense line and force line

 

 

Measurement data will include residual

 

 

as close as possible to DUT terminal.

 

 

 

 

resistance from the connection cable.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Common

 

 

 

 

 

 

 

Common

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Guard

 

 

 

 

 

 

 

 

 

Guard

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Force

 

 

 

 

 

 

 

 

 

 

 

Force

 

to

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

to

 

 

 

 

 

 

 

 

 

 

 

 

DUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Triaxial

 

 

 

 

 

 

 

 

 

Coaxial cable

Triaxial

 

 

 

 

 

 

 

 

 

Coaxial cable

 

 

 

 

Common

DUT

 

Common

 

 

connector

 

 

connector

 

 

 

 

 

 

 

 

 

 

Guard

 

 

 

 

 

 

 

 

Guard

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sense

 

 

 

 

 

 

 

 

 

 

 

Sense

Wire

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Insulator

 

 

 

 

 

 

 

 

 

 

Insulator

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Plate

 

 

 

 

 

 

 

 

 

Plate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Never connect the guard terminal to any output, including circuit common, chassis

CAUTION

 

 

 

 

 

 

ground, or the guard terminal of any other unit. Doing so may result in an

 

 

 

 

emergency condition.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

For highly accurate current forcing and measurements while minimizing leakage, surround all force and sense lines from SMU by a guard as far as possible, and make cables stable by taping.

For reducing capacitance measurement error, tape the cables to any grounded materials such as shielding box. If you use probe card, grounding the probe card also reduces capacitance measurement error.

2-14

Agilent B2200 User’s Guide, Edition 2

Page 48
Image 48
Agilent Technologies B2200A, B2201A manual Connection to the DUT Interface, Connector Guard Sense Wire Insulator Plate