VXI plug&play Driver

Driver Functions

agb220xa_openList_Q

This function will query the instrument for the channels open given in the “openchan_list”.

The parameter “openchan_list” is an array of integers with each integer representing one channel. The last number of the “openchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

The “open_status” parameter is an array of integers containing the return values of the query. The “open_status” array returned will correspond one to one with “openchan_list” parameter.

Syntax

ViStatus _VI_FUNC agb220xa_openList_Q(ViSession vi,

 

ViInt32_VI_FAR openchan_list[ ], ViInt32_VI_FAR open_status[ ] );

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

openchan_list[ ] Channel numbers to know the open status. 5 digits integer.

 

 

ABCDE. where A: card number, BC: input port number, DE:

 

 

output port number. Top zero(s) can be ignored. For example, if

 

 

A=0, BC=01, and DE=01, channel number should be 101

 

 

instead of 00101.

 

open_status[ ]

Status of the channels given in the openchan_list.

 

 

1 (opened) or 0 (closed).

agb220xa_readStatusByte_Q

This function returns the contents of the status byte register.

Syntax

ViStatus _VI_FUNC agb220xa_readStatusByte_Q(ViSession vi,

 

ViPInt16 statusByte);

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

statusByte

The contents of the status byte are returned in this parameter.

Agilent B2200 User’s Guide, Edition 2

6-29

Page 215
Image 215
Agilent Technologies B2201A, B2200A manual Agb220xaopenListQ, Agb220xareadStatusByteQ, Openstatus