VXI plug&play Driver

Driver Functions

 

port13

Couple port by the input ports 13 and 14.

 

 

0 (disable) or 1 (enable).

 

agb220xa_coupleState

 

This function controls the couple mode for the specified card.

Syntax

ViStatus _VI_FUNC agb220xa_coupleState(ViSession vi, ViInt16 cstate_cardno,

 

ViInt16 couple_state);

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

cstate_cardno

Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5

 

 

(all card) in the normal configuration mode, or 0 (all card in the

 

 

auto configuration mode). For the configuration mode, see

 

 

agb220xa_func.

 

couple_state

Couple mode. 0 (OFF) or 1 (ON).

 

agb220xa_dcl

 

This function sends a device clear (DCL) to the instrument.

 

A device clear will abort the present operation and enable the instrument to accept a

 

new command or query.

 

This is particularly useful in situations where it is not possible to determine the

 

instrument state. In this case, it is customary to send a device clear before issuing a

 

new instrument driver function. The device clear ensures that the instrument will be

 

able to begin processing the new commands.

Syntax

ViStatus _VI_FUNC agb220xa_dcl(ViSession vi);

Parameters

vi

Instrument handle returned from agb220xa_init( ).

agb220xa_detectCouplePort

This function detects the input ports connected to the Kelvin cable, and assigns them as the input couple ports that will be used for the Kelvin connection. The input couple port setting is effective for the all cards. This command overwrites the previous couple port setting.

Syntax

ViStatus _VI_FUNC agb220xa_detectCouplePort(ViSession vi);

6-20

Agilent B2200 User’s Guide, Edition 2

Page 206
Image 206
Agilent Technologies B2200A, B2201A manual Agb220xacoupleState, Agb220xadcl, Agb220xadetectCouplePort