Front Panel Operation

 

 

Setup Menus

SELF_TEST

Displays the sub menu that provides the following functions.

 

EXECUTE

 

 

Starts the controller test. Wait until PASS or FAIL is displayed. Then press the

 

 

Exit key to return to the previous menu.

 

RESULT

 

 

Displays the test result. Press the Exit key to return to the previous menu.

RELAY_TEST

Displays the sub menu that provides the following functions.

 

Before starting the relay test, disconnect cables from the input connectors. And open

 

the outputs (end of cable is OK). For the Kelvin cable, disconnect it from the output

 

connectors.

 

EXECUTE

 

 

Displays the following message.

 

 

Slot1

 

 

Press the up or down arrow key to select the module to be tested. Then press the

 

 

Enter key to start the relay test, or press the Exit key to cancel this operation.

 

 

Number of modules is 2 or more, the relay test will be automatically completed.

 

 

However, if the only one module is installed, the following message will appear.

 

 

Connect IV1 and 2. Press [Enter].

 

 

Then connect a cable between the input 1 and 2, and press Enter key.

 

 

After that, change the cable connection as shown in the displayed message and

 

 

repeat this for the other input connections (input 3 and 4, 5 and 6, 7 and 8, 9 and

 

 

10, 11 and 12, 13 and 14). Do not forget to open the connectors other than the

 

 

specified pair.

 

RESULT

 

 

Displays the following message (example). Press the up or down arrow key to

 

 

display the test result for each module. Press the Exit key to return to the

 

 

previous menu.

Slot1: Not performed yet

Agilent B2200 User’s Guide, Edition 2

3-35

Page 95
Image 95
Agilent Technologies B2201A, B2200A manual Selftest