4.4.10When S-4800 control program does not start up

When the progress bar on the login dialog window does not progress to the end, or the login dialog window disappear before completing start up, check the follows.

(1)Check if the evacuation system is running.

(2)Check if the Stage power switch is ON (for Type2 only).

(3)Check that the HV cable is put correctly to the top of the electron gun. A safety protection switch is put at the part. It is to prevent electric shock when a user removed it while HV is ON. The switch is set ON automatically when the HV cable and safety cover are put correctly . If the switch is not ON, power of HV power supply is not provided and cause troubles for starting up.

4.4.11When PC has Hanged up

When PC has hanged up while operating microscope, close windows and shut down the computer.

It is recommended to shut the POWER switch of Display unit before starting up the S-4800 program again.

(1)Press Alt, Ctrl and Delete keys simultaneously.

(2)Close Program dialog window appears. Select a program which is shown as [Not responding] on the dialog window and click the End Task button. If no program is shown as [Not responding], select a most doubtful one for hanging up. If a message [This program is not responding, It may be busy, ......] appears, click the End Task button on it.

If the computer recovered to normal state, you can start operation again. If the computer did not recovered to normal state, press Alt, Ctrl and Delete keys again. Click the Shut Down button on the Close Program dialog window. The computer will be shut down.

After a message [ Its now safe to shut your computer ] appeared, turn the power switch of the computer off.

If shutting down by Alt, Ctrl and Delete keys is not effective, turn the power switch of the computer off confirming that the hard disk access lamp is not lit.

(3)Start the computer again. If a message [Windows was not properly shut down, .......Press any key to run Scan Disk on these drives......] appeared, press C key to run Scan Disk. If Windows is started up in safe mode, shut down the computer using Start - Close Windows command on the taskbar. Then start the computer again.

4-17

Page 269
Image 269
Hitachi manual When S-4800 control program does not start up, When PC has Hanged up