Fig. 2-4 shows the Type 1 specimen stage.

[Z] control

[R] control

Absorbed current measuring terminal

[Y] control

Fig. 2-4 Specimen Stage (type I)

Fine opening control X knob Opening selector

Fine opening control Y knob [T] control

Stage lock switch

[X] control

[X]control : Moves specimen in X direction within a range from 0 to 50 mm.

[Y]control : Moves specimen in Y direction within a range from 0 to 50 mm.

[Z]control : Moves specimen in Z direction or shifts WD (Working Distance) within a range from 1.5 to 40 mm.

(X-ray analysis position is at 15 mm working distance)

[R]control : Rotates specimen through 360° (continuously variable)

[T]control : Tilts specimen within a range from -5° to 70°.

[Absorbed current measuring terminal]

: Measures beam absorbed current on specimen.

 

(Normally used as specimen applied voltage

 

terminal)

[Opening selector]

: Selects any of 4 aperture openings.

[Fine opening control X knob]

: Effects fine control in the direction for introducing or

 

taking out aperture.

[Fine opening control Y knob]

: Effects fine control in the direction orthogonal to the

 

above direction.

[Stage lock switch]

: Used for locking the specimen stage to reduce

 

vibration influence.

2 - 4