4.3Maintenance of Air CompressorOnly for P/N 539-9010 and 539-9060

NOTICE

Instruments shipped to Europe area where CE marking conformity is necessary do not provide with the air compressor. Also instruments those shall be compliant to SEMI-S2 guideline do not provide with the air compressor even when shipped to outside of Europe area.

4.3.1Checkup and Maintenance

Check the following items regularly.

Before checkup/maintenance, be sure turn the main switch off.

 

 

 

 

Service Cycle

 

Check Item

Procedure

 

 

 

 

Every

Every 250 h

Every 3,000 h

Every 6,000 h

 

 

 

 

 

 

day

(1 month)

(1 year)

(2 years)

 

 

 

 

 

 

 

Discharge of drain

Loosen the drain

 

 

 

 

 

cock of air tank to

 

 

 

 

 

discharge drain.

 

 

 

 

Activation of control

Check the

 

 

 

devices (pressure

working pressure

 

 

 

gauge, pressure switch

and activated

 

 

 

and safety valve)

function.

 

 

 

 

Abnormal vibration and

 

 

 

abnormal sound

 

 

 

Loose bolts, nuts and

If loose, retighten

 

 

 

screws

them.

 

 

 

Contamination and

Blow air to

 

clogging of strainer/filter

remove clogged

 

 

 

substance.

 

 

 

 

 

 

 

Leakage through air

 

 

 

valve

 

 

 

Piston ring

 

 

 

 

 

 

 

 

Rider ring

 

 

 

 

 

 

 

 

Bearing

Ball bearing

 

 

 

 

 

 

 

 

Needle roller

 

 

 

 

bearing

 

 

 

CAUTION

The air compressor should be drained gradually by slowly loosening the drain cocks. The inside of tank is at a high pressure. So, refrain from an abrupt cock operation, otherwise moisture may spray out to cause a danger.