INSTALLATION CONDITIONS

1.Power Requirements

100 V AC (±10%), single phase, 50/60 Hz, 4 kVA

In case to use other voltage than 100V AC, a double insulation transformer should be provided by user.

Exclusive grounding of less than 100 ohms in ground resistance should be provided by user.

2.Physical Dimensions

Main unit

S-4800I

840 (W) ×

966 (D) ×

1620 (H) mm: 583 kg

 

S-4800II

840 (W) ×

966 (D) ×

1660 (H) mm: 651 kg

Display unit

1000 (W) ×

960 (D) ×

1200 (H) mm: 205 kg

3.Installation Environment

(1)

Ambient temperature

15 °C to 25 °C (temperature variation 0.5 °C/10 min)

(2)

Relative humidity

Less than 60%, without condensation

(3)Stray magnetic field

 

 

 

 

Frequency

Frequency

 

 

 

 

 

 

Component

Component

Change with Time

 

Accelerating

Working

Magnification

Synchronous with

Asynchronous with

 

Voltage

Distance

 

Power Supply AC

Power Supply DC

(drift)

 

 

 

 

Magnetic Field

Magnetic Field

 

 

 

 

 

 

(effective value)

Fluctuation

 

 

 

15 kV

4.0 mm

220,000×

290 nT (horizontal)

320 nT (horizontal)

600 nT/min (horizontal)

Allowable

220 nT (vertical)

240 nT (vertical)

450 nT/min (vertical)

 

 

 

 

value

1 kV

1.5 mm

120,000×

150 nT (horizontal)

170 nT (horizontal)

310 nT/min (horizontal)

 

 

130 nT (vertical)

150 nT (vertical)

280 nT/min (vertical)

 

 

 

 

 

 

 

 

 

 

 

 

nT: Nano Tesla

PRECAUTION - 11

Page 43
Image 43
Hitachi S-4800 manual Power Requirements, Physical Dimensions, Installation Environment