Hitachi S-4800 manual Operation, Aperture for Each Kind of Film

Models: S-4800

1 294
Download 294 pages 5.17 Kb
Page 286
Image 286

6.1Model S-5080 Auto-Camera (option)

6.1.4Operation

(1)Installing the Camera

You should ask a service engineer for the installation. Take out the PHOTO CRT unit by referring to Figure 6-8. Screw four 5-mm screws to fix the camera on the PHOTO CRT panel.

(2)Adjustment

The adjustment for the roll film 120 is described below. Similar procedures are applicable for other kinds of film.

(a)Focusing

1)Remove the film holder, and attach the focusing hood instead.

2)Select the operation mode of NORMAL on the SEM, and press PHOTO switch.

3)Attain the optimum focus so that the raster on the focusing hood screen appears most sharp.

(b)Aperture selection

By referring to Table 6-2, select an appropriate aperture for each kind of film.

Table 6-2 Aperture for Each Kind of Film

Film

 

ISO

Aperture (f)

Polaroid type 4 x 5

 

TYPE52

400

8

Land film

 

TYPE53

800

8

 

 

 

 

 

Polaroid 107

 

 

3000

16

6 cm × 7 cm

 

SS

100

5.6

Roll film

 

TRI-X

400

8

 

 

 

 

 

NOTICE : The model S-5080 is adjusted before shipment. that readjustment of the camera is necessary, representative.

If you decide for any reason contact your nearest service

6 - 6

Page 286
Image 286
Hitachi S-4800 manual Operation, Aperture for Each Kind of Film