Fig. 2-7 shows the block diagram of evacuating system.

MV1

Electron gun

 

chamber

MV2

1st

intermediate

 

Chamber

MV3

2nd

 

intermediate

 

Chamber

 

V1

 

GV

 

SC:Specimen

 

chamber

IP1

IP2

IP3

SEC:

Specimen exchange chamber

 

 

 

 

 

 

 

PiG-2

LV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PeG

 

 

 

PiG-1

V3

leak valve

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TMP

BT V2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RP

 

 

 

 

Fig. 2-7 Evacuating System

 

IP–1, –2, –3

:

Ion pumps

 

 

 

 

 

 

TMP

:

Turbo-molecular pump

 

 

 

 

 

 

RP

:

Rotary pump

 

 

 

 

 

 

BT

:

Buffer tank

 

 

 

 

 

 

 

 

Keeps back pressure of TMP while RP is evacuating

 

 

specimen exchange chamber.

 

 

GV

: Gate valve to isolate specimen chamber and exchange chamber

MV1 to MV3

: Manual valves to evacuate gun and intermediate chambers

V1

: Valve for gun and specimen chamber isolation

V2,V3

:

Vacuum valve

 

 

 

 

 

 

PiG–1

: Pirani gauges to measure vacuum of specimen chamber

PiG–2

: Pirani gauges to measure vacuum of specimen exchange chamber

Pe

: Penning gauge to measure vacuum of specimen chamber

LV

: Leak valve to introduce air into specimen exchange chamber

RP leak valve

:

Leak valve to introduce air into rotary pump

 

2 - 7

Page 69
Image 69
Hitachi S-4800 manual Evacuating System