Hitachi S-4800 manual Setting of Accelerating Voltage and Emission Current

Models: S-4800

1 294
Download 294 pages 5.17 Kb
Page 131
Image 131

(3) Set Intensity to “2” and click Execute button.

The emission current value caused by flashing will be displayed on the Ie part of the HV Display area about 2 seconds.

For normal flashing operation, use intensity “2”. However, when the stable stage of the emission current is short, or the emission current begins to increase about 1 hour after applying HV (in that case a large quantity of gas molecules are absorbed on the cathode), use intensity “3”.

After flashing, the emission current decreases for a while, then becomes stable. (Even in the stable period, the emission current continuously decreases slightly.) The emission current becomes unstable and increases after 5 to 10 hours of operation.

Noise (bright or dark lateral stripes on the image) may appear in the initial period, but will disappear in the stable period. When there is a large amount of noise and the emission current becomes unstable, turn HV Off, and flash again.

A message will appear when 8 hours of accumulative operation time or 24 hours of total time passed after flashing, as a reminder to flash the tip again.

If the tip is not flashed within 30 minutes, the HV will be turned off automatically.

The time period can be set at other value if necessary. Contact a service engineer to change the time.

3.3.3Setting of Accelerating Voltage and Emission Current

To set the accelerating voltage and emission current, use the HV Control dialog window.

(1)To set an accelerating voltage, select a voltage from the Vacc list box. Accelerating voltage can also be set by using the keyboard to input a value followed by the Return key.

Also Up/Down buttons beside the listbox are available for 100V step setting.

(2)To set an emission current, select a current in the Set Ie to list box. It is recommended to set it at 10 µA for normal operation.

Accelerating VoltageEmission Current

Both accelerating voltage and emission current are adjusted immediately upon changing them if the HV is on.

3 - 13

Page 131
Image 131
Hitachi S-4800 manual Setting of Accelerating Voltage and Emission Current, Accelerating VoltageEmission Current