Hitachi S-4800 manual 100/200V Safety circuit

Models: S-4800

1 294
Download 294 pages 5.17 Kb
Page 62
Image 62

Use 8 square mm cable with M5 wire terminal.

CAUTION

Be sure that the output of the step down transformer is AC100V ±10%. Irregular connection causing higher voltage will result serious damage to the equipment.

Alternately the step down transformer may be put at the input of S-4800.

In the case, connect the input of the transformer to the main breaker of the site power supply. Connect the out put to the Main breaker (ELB) of S-4800. Refer fig 1-5 for the connection.

However, the former way (put the step down transformer next to the S-4800 Main breaker) is strongly recommended for safety.

100/200V

 

 

Safety circuit

 

 

MAIN-ELB

 

Current

 

 

 

100V

 

Short

Breaker

 

 

Step Down Transformer

50A

Fig 1-5 Connection of the step down transformer

In both of above cases, ground the instrument at a grounding resistance lower than 100 ohms. The grounding terminal must not be common to other electric equipment, i.e. exclusive grounding should be provided.

NOTICE

The instrument provides one of following two types of Main ELB.

(1) For using with 200 to 240V power line

Leakage current breaker (with key lock, CE/UL conforming)

Capacity

: 6.4kVA

Maximum current

: 32A

Sensitive current

: 30mA

Short-circuit breaking capacity

: 25kA

(2) For using with 100V power line

 

Leakage current breaker(without key lock, CE or UL not conforming)

Capacity

: 5kVA

Maximum current

: 50A

Sensitive current

: 30mA

Short-circuit breaking capacity

: 5kA

Instruments shipped to USA and Europe area and ordered to meet S2 guideline provide the type of above (1).

1-11

Page 62
Image 62
Hitachi S-4800 manual 100/200V Safety circuit