3.5.6Operation of the Specimen Stage (Type I - Manual Stage)

The S-4800 type I SEM provides a 5-axis manual stage.

3.5.6.1 Movable range and specimen exchange position

The following table shows movable range and specimen exchange position.

 

Movable Range

Exchange position

 

 

 

X

0 to 50 mm

25.0 mm

 

 

 

Y

0 to 50 mm

25.0 mm

 

 

 

R

0 to 360°

 

 

 

T

-5 to 70°

 

 

 

Z

1.5 to 30 mm

8.0 mm

 

 

 

3.5.6.2 Movable range limitation of X, Y and R axes

(1)When no optional detector is in its operation position, no limitation for a specimen smaller then 5 inches diameter.

For 6 inches diameter, 100mm square and 125mm square, X axis range has no limitation and Y axis range is limited as

6mm Y 50mm.

(2)When the YAG type BSE detector has been inserted to its operating position, limit X axis range as follows.

smaller than 4 inches diameter : no limitation

5 inches diameter : 2mm X 50mm

6 inches diameter : 15mm X 50mm

100mm square : 15mm X 50mm

125mm square : 15mmX 50mm

(3)For 125mm square specimen, Use rotation angle of 0,90,180 and 270 deg. only.

And rotate R axis only at X=Y=25mm (center position). If rotate specimen at other X,Y position, specimen may hit the detector.