Hitachi S-4800 manual

Models: S-4800

1 294
Download 294 pages 5.17 Kb
Page 211
Image 211

(d) Save condition

Click the Save button. The Condition Save dialog window will open.

Input comment, specify a Condition name and then, click Save button.

(2) Loading a Set of Operating Conditions

Use the following steps to load a set of operating conditions.

(a) Open the Setup dialog window

Select Condition Load from the Operate menu, or click the Setup button in the tool button section of the operation panel and select Op.Cond tab.

Select Load Memorized Condition on it.

(b) Select a condition name

Memorized condition names are listed. Select a name by clicking it.

Use sample picture and comment for reference of selection.

(c) Select items to be loaded

Check items ( Vacc to Contrast) you wish to load and reproduce. Stage position is effective only for Type2 instrument.

(d) Reproduce column condition

Click the Load button. Memorized data are retrieved and the electron optical column is set.

If the Stigma / Alignment is included, a message will appear for confirmation. These are possibly changed if the loaded condition was saved long before. In such cases, it may be better not to reproduce Stigma and Alignment data. The recent data is memorized automatically by another condition memory function (the

most recent condition for each of Vacc Condenser lens setting is memorized and reproduced automatically)

(3) Deleting an Operating Conditions

Select a condition name as same manner as above, and click the Delete button.

3-93

Page 211
Image 211
Hitachi S-4800 manual