Hitachi S-4800 manual IP1

Models: S-4800

1 294
Download 294 pages 5.17 Kb
Page 72
Image 72

The display status changes as the following table by depressing the MODE switch.

Display

Status indicator

Numeric /

description

status

lamp

character

 

 

 

display

 

Initial status

No lamp blinks

No indication

(Normal running status)

IP1

IP1 lamp blinks

IP1vacuum

XEY XY Pa

IP2

IP2 lamp blinks

IP2 vacuum

XEY XY Pa

IP3

IP3 lamp blinks

IP3 vacuum

XEY XY Pa

S.C

S.C lamp blinks

Specimen

XEY XY Pa

 

 

chamber

 

 

 

vacuum

 

S.E.C

S.E.C lamp blinks

Specimen

XEY XY Pa

 

 

exchange

 

 

 

chamber

 

 

 

vacuum

 

ECO

ECO lamp blinks

00.00

Evacuation start time

 

 

 

(refer to ECO mode setting)

BAKE

BAKE lamp blinks

10.00

Baking time

 

 

 

(refer to 4.10 Gun baking)

In the initial status, the display status changes as the following table by depressing the SELECT switch.

Display

Numeric / character display

description

status

 

 

Initial status

No indication

(Normal running status)

Date

XX.XX

month. day

Time

XX.XX

hour. minute

Cooling

FX.XXF――――

Water flow L/min

water flow

 

―――― shows water flow is off.

Other indication will appear on the Numeric / character display.

Numeric / character display

description

APXX

Waiting cooling time for the aperture heater before stopping

 

evacuation or introducing air into specimen chamber.

 

XX is in minutes.

PdXX

Waiting the TMP stops its revolution. Pd means “Power

 

Down”.

 

XX is in minutes

EXX

Alarm or error message. XX is number of the message.

POFF

“Power off”. It shows that the evacuation power (pumps,

 

valves) is shut off (control power supply is on).

2 - 10