4.Water Supply and Drain

(1)

Water flow rate

1 to 1.5 L/min

(2)

Water pressure

50 to 100 kPa

(3)

Water temperature

10 to 20 °C (temperature variation: less than 0.5 °C/10 min)

(4)

Water supply port

One port, with Rc 3/8 taper internal thread.

 

 

(If water contains significant deposit or impurities, consult

 

 

Hitachi High-Technologies Corporation separately.)

(5)

Water drain port

One port, more than 20 mm in diameter

 

 

(natural drainage type)