Hitachi S-4800 manual Line Power Requirement, Grounding

Models: S-4800

1 294
Download 294 pages 5.17 Kb
Page 53
Image 53

1.1.3Line Power Requirement

(1)Single-phase AC, 4 kVA, 50/60 Hz.

When the supplied power line voltage is 100V, the instrument works by direct connection. When the supplied power line voltage is other than 100V (200V or others), use a step down transformer to step the power line voltage down to 100V. The transformer which meets to the standard of the country shall be prepared locally.

Refer to 1.3.1 Power and Ground Wirings for connection method.

NOTICE:

(a)Line power fluctuation should be slow, and no abrupt fluctuation is allowable.

(b)Continuous energization is required

(c)Allowable line voltage fluctuation is ±10 % max.

(2)AC Power Distribution Unit should be located within 10 m of the power distribution board on the wall (since input AC cord is only 10 m long).

(3)Be sure to use an exclusive power switch on the distribution board which is provided by the user and feed the power through the switch.

1.1.4Grounding

Connect the instrument to ground based on the standard of your country.

Avoid sharing the ground terminal with other power equipments. Be sure to ground the instrument independently.

1.1.5Water Supply and Drain

The following water supply conditions should be satisfied.

Water flow rate

: 1 to 1.5 L/min

Water pressure

:

50 to 100 kPa

Water temperature

:

10 to 20 °C (temperature variation: 0.5 °C or less/10 min)

It is recommended to supply water via a cooling water circulation unit. When using city water, natural drainage is required. If an appreciable amount of deposit is found, a proper water filter should be attached. The drain port should be located at the floor level and water pressure must be free from pulsation in a short cycle.

The water temperature for cooling the objective lens must be maintained so it is not more than 7 °C lower than the room temperature. Condensation will form when the difference between the temperatures exceeds 7 °C. And if the SEM is used for a long time under the condition where condensation forms, the objective lens area will become rusted and eventually be damaged, in which case the complete set of objective lens parts will require to be replaced.

It is recommended to provide a water-leakage protection. Refer to Figure S-12 in Safety Summary section.

1-2

Page 53
Image 53
Hitachi S-4800 manual Line Power Requirement, Grounding