Hitachi S-4800 manual

Models: S-4800

1 294
Download 294 pages 5.17 Kb
Page 74
Image 74

Alarm according to gun baking

 

Message

description

recovering operation

 

No.

 

 

 

E31(+

HV cable head is not placed to t

Refer to 4.10 Gun baking.

 

buzzer)

the cable holder properly.

 

E32(+

Baking protection cover is not put.

 

 

buzzer)

 

 

 

 

 

E34(+

Ion pump high voltage is not

 

 

buzzer)

applied.

 

 

E35(+

Ion pump 1 vacuum degraded

 

 

buzzer)

while baking

 

 

E36(+

Baking time setting is not correct.

 

 

buzzer)

 

 

Alarm according to ECO mode setting

 

 

 

 

 

 

E37(+

Start gun baking while in ECO

Refer to ECO mode operation

 

buzzer)

mode

 

 

E38(+

Evacuation start time must be 5

 

 

hours or more later from present

 

 

buzzer)

 

 

time.

 

 

 

 

 

E39(+

Start ECO mode while gun baking

 

 

buzzer)

 

 

 

 

 

 

Specimen exchange rod is not at

 

 

E99(+

its proper position or at present in

 

 

buzzer)

other errors (E01, E02, E03,

 

 

 

E05, E08, E09, E06, E14)

 

Errors due to equipment failure

 

 

 

 

 

 

 

Control power supply error

Contact to service engineer

 

E01(*)

(DC24V)

 

 

 

 

 

 

E02(*)

Control power supply error

 

 

(DC15V)

 

 

 

 

 

E03(*)

Control power supply error

 

 

(DC12)

 

 

 

 

 

E04(*)

Vacuum valve did not work

 

 

correct

 

 

 

 

 

E05(*)

Rotary pump (or dry pump) is not

 

 

working

 

 

 

 

 

E06(*)

GV did not work correct

 

 

E07(*)

Filament of Pi gauge broken

 

2 - 12

Page 74
Image 74
Hitachi S-4800 manual