2.3.10SEM Data Manager Window

SEM Data Manager is an image-filing program with an easy-to-operate database function. Images are registered to the database automatically when saved.

For detail, refer to <3.12 Using SEM Data Manager>

2.3.11Alignment Dialog Window

The Alignment dialog window is used for alignment operations.

To open the dialog window, click the Wobbler button in the Scanning Image window or the

Column Alignment button on the toolbar. Or select the Alignment command from the Operate menu. When the dialog window is opened with the Wobbler button, Aperture Alignment (when the magnification mode is High Mag) or Low Mag Position (when the magnification mode is Low Mag) starts as the default setting.

Refer to <3.4.3 Column Alignment Operation>

Function select buttons

Alignment operation area

 

 

Next step button

 

 

 

 

 

 

Fig 2-26 Alignment window

1.

Function select buttons

Starts each alignment function. Off stops functions.

2.

Aperture Align button

Starts Aperture Alignment function. It is the same as the Aperture

 

 

Align select button. But it is an independent button since it is used

 

 

frequently.

 

3.

Reset button

Resets alignment of the selected function.

4.

Reset All button

Resets alignment of all functions.

5.

Alignment

This area is used for adjustment operation. When the mouse pointer

 

operation area

is

, you can make adjustment both in X or X direction by

 

 

moving the mouse pressing the left button. When the mouse pointer

 

 

is

or

, adjustment direction is restricted in X or Y direction.

6.

Close button

Stops alignment operation and closes the dialog window.

7.

Next step button

Activates next alignment function.

F4 key steps to the next alignment function, and Shift + F4 key to previous function.