CAUTION

The specimen size and height, and detectors being used must be set correctly to allow motion of each axis in the proper ranges. If this is not done, the specimen may strike the objective lens and cause damage to the specimen and objective lens

3.5.7.4Valious operation for specimen stage

The following operation methods are available.

(a)Operation on the Stage tab of the operation panel (all 5 axes)

(b)Manual operation using trackball (X and Y)

(c)Mouse operation on the image (all 5 axes)

(d)Position memory function (X, Y and R axes)

(e)Stage history function

memorize position where capturing or photographing was performed, or manually registered and display the trajectory of above points. Re-set stage to memorized position also possible

(f) Image Navigation function

go to assigned position on captured images or imported external images

(g) Rotation assist function

Assist to observe points these are possible to be observed by rotation

(h) Alignment function (option)

Enables to use sample coordinates

3 - 44

Page 162
Image 162