6-16
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
Figure 6-4 Standard Connections for a Receiver Calibration
3. To choose a non-ratioed measurement, press:
For ES analyzers, press . This sets the source at PORT 1.
4. Set any other measurement parameters that you want for the device measurement:
power, number of points, IF bandwidth.
5. To perform a receiver error correction, press:
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4, “Printing, Plotting, and Saving
Measurement Results” for procedures.
6. This completes the receiver calibration for transmission measurements. You can
connect and measure your device under test.
NOTE The accuracy of the receiver calibration will be nearly the same as the test
port power accuracy; and the test port power accuracy can be significantly
improved by performing a power mete r so u r ce calibration, as described later
in "Power Meter Measurement Calibration" on page 6-33.
Calibrations at powers other than 0 dBm are possible. Receiver calibration normalizes the
trace to the value set for the reference level. For example, to do a receiver calibration at
10 dBm, set the source to10 dBm, set the reference level to 10 dBm, then perform the
receiver calibration.
Meas
INPUT PORTS
B
TEST PORTS 1
Cal
CALIBRATE MENU
RECEIVER CAL
10 x1
TAKE RCVR CAL SWEEP