7-28
Operating Concepts
Analyzer Display Formats
Figure 7-11 Linear Magnitude Format
SWR Format
The softkey reformats a reflection measurement into its equivalent SWR (standing
wave ratio) value. See Figure 7-12. SWR is equivalent to (1 + ρ)/(1 ρ), where ρ is the
reflection coefficient. Note that the results are valid only for reflection measurements. If
the SWR format is used for measurements of S21 or S12 the results are not valid.
Figure 7-12 Typical SWR Display
SWR