6-64
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
7. After you connect the second ECal module, press to continue the
ECal.
8. Repeat steps 4 and 5 if you selected to calibrate using the manual thru option.
9. Review the displayed calibration results. Refer to "Perform the Confidence Check" on
page 6-65.
10.Save the calibration results by pressing .
Display the Module Information
Display information about the module by pressing
.
A text window is displayed that contains the following information about the selected
module.
1PC interface unit port to which it is connected 6Number of measurement points
2Model number 7Start (minimum) freq u e n cy
3Serial number 8Stop (maximum) frequency
4Connector type 9Suggested warmup time
5Date of last certification 10 Warmup status (This is the remaining time
on an internal analyzer timer. The timer
simply counts down the recommended ECal
module warm-up time. When the timer
reaches zero, a message is displayed
indicating that the ECal module is ready. The
ECal module temperature or status is not
read. The timer is reset when the analyzer is
powered up or preset or when a new module
is attached.)
CONTINUE ECal
Save/Recall
Cal
ECal MENU
CONFIGURE [ ]
MODULE INFO