1-7
Making Measurements
Measuring Magnitude and Insertion Phase Response
Measuring Magnitude and Insertion Phase Response
This measurement example shows you how to measure the maximum amplitude of a
surface acoustic wave (SAW) filter and then how to view the measurement data in the
phase format, which provides information about the phase response.

Measuring the Magnitude Response

1. Connect your test device as shown in Figure 1-2.
Figure 1-2 Device Connections for Measuring a Magnitude Response
2. Press and choose the measurement settings. For this example the
measurement parameters are set as follows:
or on ET models:
or on ET models:
You may also want to select settings for the number of data points, averaging, and IF
bandwidth.
3. Remove the device and connect the power cables together (thru) and perform a response
calibration using the following key presses.
Press .
Preset
Meas
Trans:FWD S21 ( B/R)
TRANSMISSN
Center 134 M/µ
Span 50 M/µ
Power 3x1
Scale Ref
AUTO SCALE
Chan 2
Meas
Trans:FWD S21 ( B/R)
TRANSMISSN
Scale Ref
AUTO SCALE
Chan 1 Cal
CALIBRATE MENU
RESPONSE
THRU