MODEL FCL-1056

1.3 SPECIFICATIONS — GENERAL

Sample requirements:

Pressure: 3 to 65 psig (122 to 549 kPa abs)

A check valve in the inlet prevents the sensor flow cells from going dry if sample flow is lost. The check valve opens at 3 psig (122 kPa abs). If the check valve is removed, minimum pres- sure is 1 psig (108 kpa abs).

Temperature: 32 to 122°F (0 to 50°)

Minimum Flow: 3 gal/hr (11 L/hr)

Maximum flow: 80 gal/hr (303 L/hr); high flow causes the overflow tube to back up

Sample Conductivity: >50 µS/cm at 25°C

Process connection: 1/4-in OD tubing compression fitting (can be removed and replaced with barbed fitting for soft tubing)

Drain connection: 3/4-in barbed fitting. Sample must drain to open atmosphere

Wetted parts:

Overflow sampler and flow cell: acrylic, polycar- bonate, Kynar®1, nylon, silicone

Chlorine sensor: Noryl®2, Viton®3, wood, silicone, polyethersulfone, polyester, and platinum

pH sensor (3900VP) : Stainless steel, glass, Teflon®4, polyphenylene sulfide, EPDM, and silicone

Response time to step change in chlorine concen- tration: <80 sec to 95% of final reading for inlet sample flow of 3 gph (11 L/hr)

Weight/shipping weight:

Model FCL-01: 10 lb/13 lb (4.5 kg/6.0 kg)

Model FCL-02: 11 lb/14 lb (5.0 kg/6.5 kg)

[rounded to the nearest 1 lb. (0.5 kg)]

1.4 SPECIFICATIONS — SENSOR

Free chlorine range: 0 to 10 ppm as Cl2. For higher ranges, consult the factory.

pH correction range: 6.0 to 9.5. For samples having pH between 9.5 and 10.0, consult the factory. If pH < 6.0, correction is not necessary. For manual pH correction, choose option -01. For continuous pH correction choose option -02.

Accuracy: Accuracy depends on the accuracy of the chemical test used to calibrate the sensor.

SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

Interferences: Monochloramine, permangante, peroxides.

Electrolyte volume: 25 mL (approx.)

Electrolyte life: 3 months (approx.); for best results replace electrolyte monthly.

1.5 SPECIFICATIONS — ANALYZER

Case: Polycarbonate, NEMA 4X/CSA4 (IP65)

Display: Monochromatic back-lit LCD. Main character height 0.6 in (15 mm). Display is user-programable.

Languages: English, German, Italian, Spanish, French, Portuguese

Ambient temperature and humidity: 32 to 131°F (0 to 55°C); RH 5 to 95% (con-condensing)

Storage temperature: -4 to 140°F (-20°C and 60°C)

Power: 84 to 265 Vac, 47.5-65.0 Hz, 15 W. Equipment protected by double insulation.

RFI/EMI: EN-61326

LVD: EN-61010-1

Outputs: Two 4-20 mA or 0-20 mA isolated outputs. Continuously adjustable. Linear or logarithmic. Maximum load 550 ohms. Output dampening with time constant of 5 sec is user-selectable.

Alarms: Four alarm relays. Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed with interval timer settings.

Relays: Form C, SPDT, epoxy sealed

Relay Contact ratings:

5 A at 28 VDC or 300 VAC (resistive) 1/8 HP at 120/240 VAC.

Terminal Connections Rating: Power connector (3-leads):18-12 AWG wire size. Current output connectors (2-leads):24-16 AWG wire size. Alarm relay terminal blocks: 18-16 AWG wire size

1

Kynar is a registered trademark of Elf Atochem North America.

2

Noryl is a registered trademark of General Electric.

3

Viton is a registered trademark of E.I. duPont de Nemours & Co.

4

Teflon is a registered trademark of E.I. duPont de Nemours & Co.

2

Page 12
Image 12
Emerson PN 51-FCL-1056 instruction manual Specifications General, Specifications Sensor, Specifications Analyzer